Scientech 10K Series User Manual

Browse online or download User Manual for Scales Scientech 10K Series. Scientech 10K Series User Manual [de] [fr] [it]

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1

SCIENTECH® Series 10000 Electronic Balance Setup and Operating Procedures PN10597K

Page 2 - Table of Contents:

(PCS) Front Panel Parts Counting: Note: Access to the PCS function via the front panel is disallowed if bi-directional RS-232 is enabled. RS-232 is

Page 3 - Unpacking and Setup:

(%) Percent Weighing: User Action Balance Response 1 Place an empty container on the weighing pan and press the ZERO button. Zeros are displayed. 2

Page 4 - Unlock the Clamping Screw:

Capacity Tracker: The capacity tracker provides a graphic display of the used and unused portions of the weighing range and is shown in Figure 4. Eac

Page 5 - Good Measurement Practices:

Below Balance Weighing: Note: Please use care when installing Below Balance Weighing. The suspension can be damaged if over tightened. The overload

Page 6 - Front Panel Controls:

RS-232 Interface: Note: Scientech balances are DTE (computer) devices. If communicating to another DTE device (ie computer), a Scientech RS-232 cabl

Page 7 - The ZERO Button:

User Action Balance Response 4 LOCAL CONTROL SELECTION (LC on is default). “LC on” means that the front panel switches control the balance. To en

Page 8

Using Bi-Directional Communications: Scientech electronic balances, interfaced to a computer via the RS-232 port, can be controlled by a computer prog

Page 9 - (CAL 2) Linearity:

Command Response Purpose [decimal number]_CAL[CR] None Loads a computed number based on the entered number, as a "last front panel units" r

Page 10

Type 3 Commands – Balance Output Functions: Command Response Purpose SEND[CR] See the RS-232 Balance Message Formats section. Returns the contents

Page 11 - (%) Percent Weighing:

RS-232 Balance Message Formats: The balance will respond to a SEND command by transmitting a message in one of three formats depending on how the SEND

Page 12 - Density Determination:

Thank you for choosing a Scientech Electronic Balance. Scientech, an ISO9001 registered company, and its employees are pleased to provide you with a

Page 13 - Options:

Format B - Balance in normal weighing mode: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 X X X X X X X _ _ _ Y Y Y _ _ Y Y ...CR LF Display sh

Page 14 - RS-232 Interface:

Troubleshooting Guide: Note: Many problems such as the balance remaining in the turn on sequence, "weird" characters on the display or &qu

Page 15

Environmental Requirements: This product is intended for indoor use at altitudes up to 2000 meters, Pollution Degree 1 or 2 in accordance with IEC 664

Page 16

Specifications for SA Analytical Balances: Model SA 80 SA 120 SA 210 SA 310 SA 410 SA 510 SA 210D SA 525D Weighing Mode Single Range Dual Ra

Page 17

Specifications for SP Precision Balances: Model SP150 SP250 SP350 SP500 SP1000 SP1500 SP404D SP510D SP2050DWeighing Mode Single Ran

Page 18

Specifications for SL Laboratory Balances: Model SL400 SL600 SL1000 SL2000 SL3000 SL5000 SL6000 SL3100D SL5200DWeighing Mode Single

Page 19

Specifications for SG General Balances and SE Gem Balances: Model SG3000 SG5000 SG8000 SG12000 SE250 SE2500Weighing Mode Single Range Cap

Page 20 - Format C:

Limited Warranty: All Scientech Series 10000 Electronic Balances and their accessories are warranted against defects in materials and workmanship with

Page 21 - Troubleshooting Guide:

Quick Setup Guide: Adjust the FRONT FEET to level the balance by centering the LEVEL INDICATOR.5 CONNECT BALANCE TO ELECTRICAL OUTLET3 LEVEL BALAN

Page 22 - Environmental Requirements:

Environmental Requirements:...

Page 23

Choose the Proper Environment: The environment in which your balance is used is very important. Air movement, temperature changes, vibrations, direct

Page 24

Connect to an Electrical Outlet: The balance is supplied with a 115/220, 50/60 Hz automatic switching, remote power supply. This power supply will wo

Page 25

Environment: 1. The more stable your environment, the better your measurement results. Changes in temperature, pressure, and humidity affect balance

Page 26

The RANGE/FUNCTION Button: Single Range Balance Models: Pressing the RANGE/FUNCTION button begins the following menus cycle: • PCS – Front panel part

Page 27 - Returned Goods Procedure:

Balance Functions and Features: Autocalibration – To Calibrate the balance to an External Weight: This procedure calibrates your balance to an externa

Page 28 - Quick Setup Guide:

Autocalibration – To Calibrate the Balance and its Motorized Internal Calibration Weight to an External Weight: This procedure is only for SA and SM b

Comments to this Manuals

No comments